TDS2012C
|
TDS2014C
|
TDS2022C
|
TDS2024C
|
Display (QVGA LCD)
|
TFT
|
TFT
|
TFT
|
TFT
|
TFT
|
TFT
|
TFT
|
Bandwidth*3
|
50 MHz
|
70 MHz
|
70 MHz
|
100 MHz
|
100 MHz
|
200 MHz
|
200 MHz
|
Channels
|
2
|
2
|
4
|
2
|
4
|
2
|
4
|
External Trigger Input
|
Included on all models
|
Sample Rate on Each Channel
|
500 MS/s
|
1.0 GS/s
|
1.0 GS/s
|
2.0 GS/s
|
2.0 GS/s
|
2.0 GS/s
|
2.0 GS/s
|
Record Length
|
2.5k points at all time bases on all models
|
Vertical Resolution
|
8 bits
|
Vertical Sensitivity
|
2 mV to 5 V/div on all models with calibrated fine adjustment
|
DC Vertical Accuracy
|
±3% on all models
|
Vertical Zoom
|
Vertically expand or compress a live or stopped waveform
|
Maximum Input Voltage
|
300 VRMS CAT II; derated at 20 dB/decade above 100 kHz to 13 Vp-p AC at 3 MHz
|
Position Range
|
2 mV to 200 mV/div +2 V
>200 mV to 5 V/div +50 V
|
Bandwidth Limit
|
20 MHz for all models
|
Input Coupling
|
AC, DC, GND on all models
|
Input Impedance
|
1 MΩ in parallel with 20 pF
|
Time Base Range
|
5 ns to 50 s/div
|
5 ns to 50 s/div
|
5 ns to 50 s/div
|
2.5 ns to 50 s/div
|
2.5 ns to 50 s/div
|
2.5 ns to 50 s/div
|
2.5 ns to 50 s/div
|
Time Base Accuracy
|
50 ppm
|
Horizontal Zoom
|
Horizontally expand or compress a live or stopped waveform
|
I/O Interfaces
|
USB Ports
|
USB host port on front panel supports USB flash drives
USB device port on back of instrument supports connection to PC and all PictBridge-compatible printers
|
GPIB
|
Optional
|
Nonvolatile Storage
|
Reference Waveform Display
|
(2) 2.5k point reference waveforms
|
Waveform Storage without USB Flash Drive
|
(2) 2.5k point
|
(2) 2.5k point
|
(4) 2.5k point
|
(2) 2.5k point
|
(4) 2.5k point
|
(2) 2.5k point
|
(4) 2.5k point
|
Maximum USB Flash Drive Size
|
64 GB
|
Waveform Storage with USB Flash Drive
|
96 or more reference waveforms per 8 MB
|
Setups without USB Flash Drive
|
10 front-panel setups
|
Setups with USB Flash Drive
|
4000 or more front-panel setups per 8 MB
|
Screen Images with USB Flash Drive
|
128 or more screen images per 8 MB (the number of images depends on file format selected)
|
Save All with USB Flash Drive
|
12 or more Save All operations per 8 MB
A single Save All operation creates 3 to 9 files (setup, image, plus one file for each displayed waveform)
|
*3 Bandwidth is 20 MHz at 2 mV/div, all models.
Acquisition Modes
Mode
|
Description
|
Peak Detect
|
High-frequency and random glitch capture. Captures glitches as narrow as 12 ns (typical) at all time base settings from 5 μs/div to 50 s/div
|
Sample
|
Sample data only
|
Average
|
Waveform averaged, selectable: 4, 16, 64, 128
|
Single Sequence
|
Use the Single Sequence button to capture a single triggered acquisition sequence
|
Roll
|
At acquisition time base settings of >100 ms/div
|
Trigger System
Characteristic
|
Description
|
Trigger Modes
|
Auto, Normal, Single Sequence
|
Trigger Types
Trigger
|
Description
|
Edge (Rising/Falling)
|
Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF Reject, LF Reject
|
Video
|
Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM)
|
Pulse Width (or glitch)
|
Trigger on a pulse width less than, greater than, equal to, or not equal to, a selectable time limit ranging from 33 ns to 10 s
|
Trigger Source
Characteristic
|
Description
|
2-channel Models
|
CH1, CH2, Ext, Ext/5, AC Line
|
4-channel Models
|
CH1, CH2, CH3, CH4, Ext, Ext/5, AC Line
|
Trigger View
Displays trigger signal while Trigger View button is depressed.
|
Trigger Signal Frequency Readout
Provides a frequency readout of the trigger source.
|
Cursors
Characteristic
|
Description
|
Types
|
Amplitude, Time
|
Measurements
|
ΔT, 1/ΔT, ΔV
|
Automatic Waveform Measurements
Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, RMS, Cycle RMS, Cursor RMS, Duty Cycle, Phase, Delay
|
Waveform Math
Characteristic
|
Description
|
Operators
|
Add, Subtract, Multiply, FFT
|
FFT
|
Windows: Hanning, Flat Top, Rectangular
2048 sample points
|
Sources
|
2-channel models
|
CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2
|
4-channel models
|
CH1 – CH2, CH2 – CH1, CH3 – CH4, CH4 – CH3, CH1 + CH2, CH3 + CH4, CH1 × CH2, CH3 × CH4
|
Autoset Menu
Signal Type
|
Autoset Menu Choices
|
Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo Autoset.
|
Square Wave
|
Single Cycle, Multicycle, Rising or Falling Edge
|
Sine Wave
|
Single Cycle, Multicycle, FFT Spectrum
|
Video (NTSC, PAL, SECAM)
|
Field: All, Odd, or Even
Line: All or Selectable Line Number
|
Autorange
Automatically adjust vertical and/or horizontal oscilloscope settings when probe is moved from point to point, or when the signal exhibits large changes.
|
Display Characteristics
Characteristic
|
Description
|
Display
|
QVGA Active Color TFT
|
Interpolation
|
Sin(x)/x
|
Display Types
|
Dots, vectors
|
Persistence
|
Off, 1 s, 2 s, 5 s, infinite
|
Format
|
YT and XY
|
Multiple-language User Interface and Context-sensitive Help
Characteristic
|
Description
|
Languages Available
|
English, French, German, Italian, Japanese, Korean, Portuguese, Russian*4, Simplified Chinese, Spanish, Traditional Chinese
|
*4 Requires Russian firmware, indicated by "RUS" suffix.
Environmental and Safety
Characteristic
|
Description
|
Temperature
|
Operating
|
0 to +50 °C
|
Nonoperating
|
–40 to +71 °C
|
Humidity
|
Operating and nonoperating
|
Up to 80% RH at or below +40 °C
Up to 45% RH up to +50 °C
|
Altitude
|
Operating and nonoperating
|
Up to 3,000 m
|
Electromagnetic Compatibility
|
Meets Directive 2004/108/EC, EN 61326-2-1 Class A; Australian EMC Framework
|
Safety
|
UL61010-1:2004, CSA22.2 No. 61010-1:2004, EN61010-1:2001, IEC61010-1:2001
|
Physical Characteristics
Instrument
|
Dimensions
|
mm
|
in.
|
Width
|
326.3
|
12.85
|
Height
|
158.0
|
6.22
|
Depth
|
124.2
|
4.89
|
Weight
|
kg
|
lb.
|
Instrument Only
|
2.0
|
4.4
|
With accessories
|
2.2
|
4.9
|
Instrument Shipping
|
Package Dimensions
|
mm
|
|
|
|